Combined synopsis and solicitation, first published as solicitation
Electrosurgical unit (ESU)
75F40126Q00277
Food and Drug Administration, FDA Office of Acq Grant SVCS. Surgical and Medical Instrument Manufacturing.
Response deadline
September 16, 2026 at 1:00 PM EDT
Closed 47 days ago. Posted September 10, 2026, first published August 28, 2026. Scheduled to archive September 17, 2026.
Get the next one like this from Food and Drug Administration by email, the day it is posted.Find my opportunities
This one closed 47 days ago
17 like it are open now in surgical and medical instrument manufacturing
- NX EQ EYE LANE (VA-27-00010051)
Department of Veterans Affairs, Sac Frederick
PresolicitationSDVOSBNAICS 339112Anywhere in the US36C10X27Q0016Closes in 3 daysNov 5 - Intent to Sole Source - Procurement of BD Mission™ Disposable Core Biopsy Instrument (semi-automatic, single-use)
National Institutes of Health, National Institutes of Health - CC
PresolicitationNAICS 339112Anywhere in the USNOI-OLAO-CC-27-000988Closes in 6 daysNov 8 - Anesthesia Units Amendment 4
Department of Veterans Affairs, 241-Network Contract Office 01
PresolicitationNAICS 339112Anywhere in the US36C24125R0116Closes todayNov 2
All 17 open in surgical and medical instrument manufacturing
Description
As published on SAM.gov.
All equipment shall be newly manufactured. It shall not be used, refurbished, reconditioned, remanufactured, or previously used for demonstration. The equipment and all included accessories shall be new, unused, and in full working condition at the time of delivery.
Salient characteristics and any single-source need
The salient physical, functional, and performance characteristics the equipment must meet are stated under Minimum technical requirements below:
Minimum technical requirements
The equipment shall meet the following minimum requirements.
Display and user interface
-
A touchscreen display for user interaction and parameter control
Tissue impedance monitoring
-
Monitoring of tissue impedance at a minimum rate of 434,000 times per second, providing real-time feedback and energy delivery adjustment
Output frequency
-
Output frequency within the range of 400 kHz to 500 kHz
Operating modes
The equipment shall support the following operating modes.
-
Monopolar CUT (Pure)
-
Monopolar CUT (Blend)
-
Monopolar COAG (Fulgurate)
-
Monopolar COAG (Spray)
-
Monopolar COAG (Soft)
-
Bipolar coagulation
-
Vessel sealing
Electrical and safety
-
Compliance with all applicable Federal, state, and local safety standards
Required accessories
The contractor shall provide all accessories necessary to fully operate the equipment in every operating mode listed above. At minimum the following shall be included.
-
Monopolar electrode and holster with switch, compatible with all monopolar modes.
-
Patient return electrode for monopolar operation.
-
Bipolar forceps, cord, and footswitch for bipolar coagulation.
-
Vessel sealer with switch for vessel sealing mode.
All accessories shall be compatible with the delivered equipment and ready for use upon delivery.
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- Attachment 1 Requirements_Document_ESU_CDRH-2026-134011_jwg.docx37 KB, September 10, 2026
- Attachment 1 Requirements_Document_ESU_CDRH-2026-134011_jwg.docx37 KB, August 28, 2026
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
August 28, 2026
Solicitation
Due September 2, 2026 at 12:00 PM EDT. SAM.gov, notice 77fbe29ad43b4dfbbdc57226d07f9fe9
1 revisions on SAM
- August 28, 2026: publish, solicitation
September 10, 2026
Combined synopsis and solicitation
Due September 16, 2026 at 1:00 PM EDT. SAM.gov, notice 93cb831e158f4ae090cfb682c880d645
1 revisions on SAM
- September 10, 2026: publish, combined synopsis and solicitation
Points of contact
- Reagan Davisreagan.davis@fda.hhs.gov3017960910