# Electrosurgical unit (ESU)

Canonical: https://abierto.us/opportunities/75f40126q00277

- Solicitation number: 75F40126Q00277
- Notice type: Combined synopsis and solicitation (first published as solicitation)
- Status: Open. Offers due September 16, 2026 at 1:00 PM EDT
- Department: Department of Health and Human Services
- Agency: Food and Drug Administration
- Contracting office: FDA Office of Acq Grant SVCS (75F401)
- NAICS: 339112 Surgical and Medical Instrument Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Place of performance: Silver Spring, Maryland
- First posted: August 28, 2026
- Last posted: September 10, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/93cb831e158f4ae090cfb682c880d645/view

## Description

All equipment shall be newly manufactured. It shall not be used, refurbished, reconditioned, remanufactured, or previously used for demonstration. The equipment and all included accessories shall be new, unused, and in full working condition at the time of delivery.

Salient characteristics and any single-source need

The salient physical, functional, and performance characteristics the equipment must meet are stated under Minimum technical requirements below:

Minimum technical requirements

The equipment shall meet the following minimum requirements.

Display and user interface

	A touchscreen display for user interaction and parameter control

Tissue impedance monitoring

	Monitoring of tissue impedance at a minimum rate of 434,000 times per second, providing real-time feedback and energy delivery adjustment

Output frequency

	Output frequency within the range of 400 kHz to 500 kHz

Operating modes

The equipment shall support the following operating modes.

	Monopolar CUT (Pure)

	Monopolar CUT (Blend)

	Monopolar COAG (Fulgurate)

	Monopolar COAG (Spray)

	Monopolar COAG (Soft)

	Bipolar coagulation

	Vessel sealing

Electrical and safety

	Compliance with all applicable Federal, state, and local safety standards

Required accessories

The contractor shall provide all accessories necessary to fully operate the equipment in every operating mode listed above. At minimum the following shall be included.

	Monopolar electrode and holster with switch, compatible with all monopolar modes.

	Patient return electrode for monopolar operation.

	Bipolar forceps, cord, and footswitch for bipolar coagulation.

	Vessel sealer with switch for vessel sealing mode.

All accessories shall be compatible with the delivered equipment and ready for use upon delivery.

## Attachments

- [Attachment 1 Requirements_Document_ESU_CDRH-2026-134011_jwg.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/2b7dd373efd9482b9f1d60c82f38cb85/download)

## Publications

- August 28, 2026: Solicitation, due September 2, 2026 at 12:00 PM EDT. Notice 77fbe29ad43b4dfbbdc57226d07f9fe9. https://sam.gov/workspace/contract/opp/77fbe29ad43b4dfbbdc57226d07f9fe9/view
- September 10, 2026: Combined synopsis and solicitation, due September 16, 2026 at 1:00 PM EDT. Notice 93cb831e158f4ae090cfb682c880d645. https://sam.gov/workspace/contract/opp/93cb831e158f4ae090cfb682c880d645/view

## Points of contact

- Reagan Davis, reagan.davis@fda.hhs.gov, 3017960910

---