Walk-through Ferromagnetic Dual Detector System

36C25524Q0433

Department of Veterans Affairs, 255-Network Contract Office 15. Electromedical and Electrotherapeutic Apparatus Manufacturing.

Awarded

Meak Solutions LLC

$34,998.00 obligated so far on USAspending

Description

As published on SAM.gov.

Statement of Requirement Replacement including installation of the walk-through ferromagnetic dual detector system with equivalent system for notification of MR projectile hazards. Ferromagnetic detection system must have adjustable acoustic and multiple location optical alarm capabilities with technology designed to decrease alarm fatigue and false- positive notifications. System shall have operator programable sensitivity limitation settings.

Detector(s) able to be mounted to entrance doorway of the current design for the magnet room (zone 3). Consideration of structural factors accordant for product include ability to mount on outside of cased opening with a seven (7) inch inset with permanent hardwired sign located eight (8) inches above door frame. The product shall be compatible for door with outswing operation only. See Appendix Basis of Design Product design consistent to existing position of power supply at 14 inches from framed casing of door.

Must have standard AC power with no special requirements. Detector range must accommodate the seventy-two (72) inch high by fifty-three (53) inch wide coverage area. Designed to withstand potential impact of hospital equipment during patient transfers with MR safe equipment. On-site training for staff shall be provided for safe operation, problem-solving, and information on technical support.

System must be ferromagnetic detector system designed to alarm on only ferromagnetic objects, shall not accept a metal detector designed to alarm on both ferrous and non- ferrous metal types. Complies with IEC 60601-1-8 AMD2:2020

Conforms to IEC 60101-1:2005 medical device specification requirements (the Ferromagnetic detectors are not considered to be medical devices; however, they are an integral part of the medical environment. As such, it is highly important that the power supply used in the detectors complies with standards for medical equipment used by the FDA, UL, as well as agencies around the world).

The contract, on USAspending

Federal procurement data the awarding office reported to FPDS, matched to this solicitation by its number.

UEI
XKHXVFCXS4R9
CAGE
78FU8
Vendor location
Mentor, OH
Contract
36C25524P0525, purchase order
Obligated
$34,998.00
Actions
1 between July 22, 2024 and July 22, 2024
Competition
Competed Under SAP, 1 offer received
Set-aside reported
Service Disabled Veteran Owned Small Business SET-ASIDE
Described as
Foreign Body Detector
Match
solicitation number 36C25524Q0433 equals the FPDS solicitation identifier; same awarding office 36C255 (high confidence)

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. June 10, 2024

    Solicitation

    Due June 17, 2024 at 12:00 PM EDT. SAM.gov, notice 5f718355173d4da3a4efc69428342865

Points of contact