Amendment 1 to 7 Ultrasonic Flowmeters

1333ND26QNB610465

National Institute of Standards and Technology, Department of Commerce NIST. Analytical Laboratory Instrument Manufacturing.

Response deadline

September 4, 2026 at 12:00 PM EDT

Closed 12 days ago. Posted August 28, 2026, first published August 21, 2026. Scheduled to archive September 19, 2026.

Description

As published on SAM.gov.

Amendment 1to RFQ 1333ND26QNB610465 Clarification questions:

Vendor Question 1: Could you please confirm the solids content (TSS) of the water? If this is clean water, with less than 2% TSS, please let us know so we can determine the appropriate unit for the application.

NIST Response: This water is moderately clean, not a slurry (2% TSS is closer to wastewater) but not filtered Vendor Question 2: The factory has confirmed that no software is required to configure the meter, as very few ultrasonic meters require software for configuration.

NIST Response: PC-based data logging is a nice-to-have but not a necessity.

Vendor Question 3: (clarification on the following process conditions for the primary and secondary cooling systems) Operating temperature of the process fluid (or pipe surface temperature) at the measurement points, for both the primary and secondary cooling loops. This determines whether standard or high-temperature transducers are required.

NIST Response: No higher than 114 degrees Fahrenheit.

Vendor Question 4: (clarification on the following process conditions for the primary and secondary cooling systems) Operating pressure of the cooling water at the measurement points (primary and secondary).

NIST Response: No higher than 10 psia.

Vendor Question 5: (clarification on the following process conditions for the primary and secondary cooling systems) Confirmation of the process fluid — the SOW references cooling water; please confirm it is treated water with no significant entrained solids or aeration that could affect ultrasonic signal transmission.

NIST Response: Primary – Heavy water, purified. Secondary - Water with minimal solids. Some algae to be expected but not significant. As defined in the attached Combined Synopsis Solicitation, Statement of Work and RFO Provisions and Clauses.

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. August 21, 2026

    Combined synopsis and solicitation

    7 Ultrasonic Flowmeters

    Due September 4, 2026 at 12:00 PM EDT. SAM.gov, notice f8d0e02996134321bf21626c4939dc3b

  2. August 28, 2026

    Combined synopsis and solicitation

    Due September 4, 2026 at 12:00 PM EDT. SAM.gov, notice c57fd080570247528e4049abe26cec79

Points of contact