# Amendment 1 to 7 Ultrasonic Flowmeters

Canonical: https://abierto.us/opportunities/1333nd26qnb610465

- Solicitation number: 1333ND26QNB610465
- Notice type: Combined synopsis and solicitation
- Status: Closed. Deadline was September 4, 2026 at 12:00 PM EDT
- Department: Department of Commerce
- Agency: National Institute of Standards and Technology
- Contracting office: Department of Commerce NIST (1333ND)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6640 Laboratory Equipment and Supplies
- Place of performance: Gaithersburg, Maryland
- First posted: August 21, 2026
- Last posted: August 28, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/c57fd080570247528e4049abe26cec79/view

## Description

Amendment 1to RFQ 1333ND26QNB610465 Clarification questions:

**Vendor Question 1:** Could you please confirm the solids content (TSS) of the water? If this is clean water, with less than 2% TSS, please let us know so we can determine the appropriate unit for the application.

**NIST Response:** This water is moderately clean, not a slurry (2% TSS is closer to wastewater) but not filtered Vendor Question 2: The factory has confirmed that no software is required to configure the meter, as very few ultrasonic meters require software for configuration.

**NIST Response:** PC-based data logging is a nice-to-have but not a necessity.

**Vendor Question 3:** (clarification on the following process conditions for the primary and secondary cooling systems) Operating temperature of the process fluid (or pipe surface temperature) at the measurement points, for both the primary and secondary cooling loops. This determines whether standard or high-temperature transducers are required.

**NIST Response:** No higher than 114 degrees Fahrenheit.

**Vendor Question 4:** (clarification on the following process conditions for the primary and secondary cooling systems) Operating pressure of the cooling water at the measurement points (primary and secondary).

**NIST Response:** No higher than 10 psia.

**Vendor Question 5:** (clarification on the following process conditions for the primary and secondary cooling systems) Confirmation of the process fluid — the SOW references cooling water; please confirm it is treated water with no significant entrained solids or aeration that could affect ultrasonic signal transmission.

**NIST Response:** Primary – Heavy water, purified. Secondary - Water with minimal solids. Some algae to be expected but not significant. As defined in the attached Combined Synopsis Solicitation, Statement of Work and RFO Provisions and Clauses.

## Publications

- August 21, 2026: Combined synopsis and solicitation, due September 4, 2026 at 12:00 PM EDT. Notice f8d0e02996134321bf21626c4939dc3b. https://sam.gov/workspace/contract/opp/f8d0e02996134321bf21626c4939dc3b/view
- August 28, 2026: Combined synopsis and solicitation, due September 4, 2026 at 12:00 PM EDT. Notice c57fd080570247528e4049abe26cec79. https://sam.gov/workspace/contract/opp/c57fd080570247528e4049abe26cec79/view