Automated Liquid Handling Platform
75F40126Q00423
Food and Drug Administration, FDA Office of Acq Grant SVCS. Analytical Laboratory Instrument Manufacturing.
Response deadline
September 23, 2026 at 11:00 AM EDT
Closed 38 days ago. Posted September 22, 2026, first published September 8, 2026. Scheduled to archive September 24, 2026.
Get the next one like this from Food and Drug Administration by email, the day it is posted.Find my opportunities
This one closed 38 days ago
39 like it are open now in analytical laboratory instrument manufacturing
- Graphite Furnace Atomic Absorption Spectrometer
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Anywhere in the USNIST-MML-FY27-SS01Closes in 28 daysNov 28 - Sources Sought Notice for Mid-IR Optical Frequency Comb
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Anywhere in the USNIST-SS26-CHIPS-13Closes in 33 daysDec 3 - Sources Sought Notice for Continuous Wave Parametric Oscillator (CW OPO) Laser System
National Institute of Standards and Technology, Department of Commerce NIST
Sources soughtNAICS 334516Anywhere in the USNIST-SS26-CHIPS-16Closes in 35 daysDec 5
All 39 open in analytical laboratory instrument manufacturing
Description
As published on SAM.gov.
Ammendment 0001: The response due date has been extended until 11am Wednesday September 23, 2026. After that time no further responses will be considered.
All equipment shall be newly manufactured. It shall not be used, refurbished, reconditioned, remanufactured, or previously used for demonstration.
Salient characteristics and any single-source need
The equipment shall meet the following salient characteristics: a benchtop automated liquid handling platform, user-friendly and easy to operate and clean, of space-saving and energy-saving design, capable of fast automatic pipetting of multiple solutions and of serial dilutions across or down the plate; able to accommodate variable liquid volumes; able to pipette into single rows, single columns, or individual wells as selected by the user in a 96-well or 384-well format, with the well format interchangeable based on user requirements; customizable to the laboratory's assay needs; equipped with programmable operating temperatures permitting use across a range of temperatures; and usable within the containment enclosure identified in the minimum technical requirements below to permit handling of sterile or hazardous materials.
Based on our research, only a specific make and model meets our needs, the Agilent Technologies Bravo Automated Liquid Handling Platform. It is the only equipment in its class that can meet our minimum needs as listed in the specifications. Other machines do not dispense low volumes, are much larger than this machine and do not fit within a biosafety cabinet.
The equipment shall meet the following minimum requirements:
-
The platform shall be benchtop in format and shall fit and operate within a biosafety cabinet of dimensions of 46in W X 24in D X 30in H that will support 120 lbs without disrupting the enclosure airflow.
-
The platform shall perform serial dilutions either across the plate or down the plate using standard disposable tip heads.
-
The platform shall use 96-channel or 384-channel pipetting heads that are interchangeable, with independent tip pickup capability.
-
The platform shall provide a minimum 15-position deck capacity.
-
The platform shall dispense from 0.1 to 200 microliters. Dispensing accuracy and precision shall be no worse than 5% deviation.
-
The platform shall be configurable to support shaking, heating, cooling, and filtration. These configurations are not required to be delivered with the platform; however, the platform shall be capable of accepting these configurations at a later date, if desired.
-
The platform shall include a controller and software to generate custom protocols.
-
The platform shall include a dedicated computer to control the machine.
-
The platform shall be compatible with standard format microplates (96-well and 384-well).
-
The platform shall operate under ambient conditions of 5 to 40 degrees Celsius and 10 to 90 percent relative humidity, non-condensing. No deck or plate temperature-control range is required.
-
Compliance with all applicable Federal, state, and local safety standards.
Attachments
Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.
- Requirements_Automated_Liquid_Handler.docx36 KB, September 8, 2026
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
September 8, 2026
Solicitation
Due September 11, 2026 at 12:00 PM EDT. SAM.gov, notice 2aea491d0bd946f68af4c7b47df91c28
2 revisions on SAM
- September 8, 2026: publish, solicitation
- September 8, 2026: publish, solicitation
September 8, 2026
Solicitation
Due September 11, 2026 at 12:00 PM EDT. SAM.gov, notice b5cd97229b4946d89b8a96110342f6fa
2 revisions on SAM
- September 8, 2026: publish, solicitation
- September 8, 2026: publish, solicitation
September 17, 2026
Presolicitation
Due September 22, 2026 at 11:00 AM EDT. SAM.gov, notice e79c67536d424aa69fda50e8ad1ecf0d
1 revisions on SAM
- September 17, 2026: publish, presolicitation
September 22, 2026
Presolicitation
Due September 23, 2026 at 11:00 AM EDT. SAM.gov, notice 80474b69df884b6c9d57ba9b0210fa5e
2 revisions on SAM
- September 22, 2026: publish, presolicitation
- September 17, 2026: publish, presolicitation
Points of contact
- Reagan Davisreagan.davis@fda.hhs.gov3017960910