SCOPE I. Minimum Performance Requirements/Salient Characteristics Salient Characteristics The contractor shall provide all resources necessary to accomplish the tasks and deliverables described in this statement of work, including the hardware system requirements which are the minimum requirements. Equivalent requirements that differ from these minimum requirements must be justified by the proposing vendor and evaluated by the NFCC prior to purchase.
1.0 General Requirement The National Forensic Chemistry Center (NFCC) requires a Bruker Alicona InfiniteFocus G6 (IFM G6) optical 3D measurement system. This non-contact, optical 3D imaging and measurement system will be used by NFCC's Trace Examination Section (TES) for surface profilometry and difference analysis measurements, including tablet debossing, punch face embossing, surface abrasion, pit measurements, corroded surface analysis, and pharmaceutical packaging topography and print profiling.
The system must expand analytical throughput to meet increasing demand for forensic 3D image analysis from law enforcement stakeholders. 2.0 Core Technology and Measurement Principle 2.1 The system shall be based on the principle of Focus Variation Microscopy (FVM), utilizing Bruker Alicona's proprietary Focus Variation algorithms to reconstruct high-resolution 3D surfaces.
2.2 The system must include advanced illumination technology (e.g., ring light, coaxial light) to enable accurate measurement of surfaces with steep angles, up to and exceeding 85 degrees. 2.3 The system must be capable of measuring both form and surface roughness parameters in a single measurement, with all data being co-registered. 2.4 The system must have the ability to measure highly reflective materials without surface treatment.
The system must be capable of acquiring accurate 3D topographic data on reflective and specular materials without requiring surface alteration that could compromise the integrity of forensic samples. 2.5 The system must have the capability to capture true-color optical data co-registered with 3D topography. This capability supports surface characterization, documentation, and interpretation of coating or material variations. Many profilometry systems such as systems that use laser or tip technologies do not provide integrated true-color 3D datasets.
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
September 8, 2026
Combined synopsis and solicitation
Due September 14, 2026 at 3:00 AM EDT. SAM.gov, notice 44430997898d4b699da81c4f883899c2
Points of contact
- Pamela T. Leepamela.lee@fda.hhs.gov2404023847
- Madeline Bryantmadeline.bryant@fda.hhs.gov2404027611
Also open from this buyer
- Electrosurgical unit (ESU)Combined synopsis and solicitationNAICS 339112Maryland75F40126Q00277Closes todaySep 16
- Endoscopy SystemsCombined synopsis and solicitationNAICS 423490Maryland75F40126Q00381Closes todaySep 16
- QELS Dynamic Light Scattering DetectorCombined synopsis and solicitationNAICS 334516Missouri75F40126Q00371Closes todaySep 16
- Human Hindbrain Organoid Functional Assay Services.Combined synopsis and solicitationNAICS 541714RFQ_CDR-2026-135968Closes tomorrowSep 17
- Walk-In Laboratory CoolerSolicitationNAICS 33341575F40126Q00386Closes in 2 daysSep 18