Skip to content
Abierto

Solicitation, first published as sources sought

WATER PURFICATION SYSTEM HCC1 LAB

36C24624Q0414

Department of Veterans Affairs, 246-Network Contracting Office 6. Commercial and Service Industry Machinery Manufacturing.

Awarded

Franklin Young International, Incorporated

$31,455.00 obligated so far on USAspending

Description

As published on SAM.gov.

VA Chesapeake Community Based Outpatient Clinic (CBOC) Chesapeake HCC1 Salient Characteristics Water Purification System that is capable of producing not less than 200 liters of water per hour, fitted with deionization capabilities MODEL MEDICA-R 200 Make-up rate @15°C 200 l/hr Delivery flow rate maximum115V, 60Hz 21 l/min @ 45psi (3bar) Inorganics (resistivity @25°C) >10 M -cm (1) Organics (TOC) - typical <30ppb -500 PPB (1) Bacteria - typical <10 CFU/ml (2) Particles 0.2 m filter Silica <0.05 mg/l Notes: 1 with Ion-exchange cylinder installed (Nuclear or Hypex grade resin), 2 subject to correct operating and maintenance procedures Feedwater Requirement:

Source: Tap water as detailed below Conductivity: <1000 S/cm Contaminant:

Measure:

Range:

Pretreatment:

Calcium: Ca ppm as CaCO3 <250 None3 >250 Softener or use very low RO recovery3 Free chlorine: Cl2ppm <0.1 None 0.1 0.5 20 inch carbon block >0.5 Cylinder of carbon sized correctly to obtain <0.1 ppm Chloramine: Cl2 ppm <0.1 None 0.1 0.5 20 inch high efficiency carbon (5) >0.5 Cylinder of carbon sized correctly to obtain <0.1 ppm Silica: SiO2 ppm <30 None >30 20 inch cartridge depth filter or use very low RO recovery (4) Fouling index: FI <10 None 10 to 20 20 inch cartridge depth filter >20 Back washable media filter with a minimum flow rate of 20 l/min Iron/manganese: Fe/Mn ppm <0.05 None >0.05 20 inch cartridge depth filter (5) >0.1 Back-washable Fe filter (6) Organics: TOC ppm C <2 None 2 to 3 20 inch carbon block (4) - >3 Cylinder of carbon sized correctly for TOC demand (4) Temperature: 4 - 40°C (Recommended 15 - 25°C) - Flowrate: (requirement at 15°C) 20 l/min Drain requirements: (gravity fall with air gap) 45 l/min Feedwater pressure: 60psi (4bar) maximum, 30psi (2bar) minimum Notes: 3) Check LSI, increase frequency of acid cleaning, 4) Increase frequency of alkaline cleaning, 5) Please use ELGA Part Numbers LA683 (Filter Housing) and LC172 (Carbon Filter), 6) Increase frequency of acid cleaning Wall mount canister filter will need (Wall backing required) Must include:

Installation, Commissioning, Training Electrical Requirements Mains Input: 115V, 60Hz System control voltage (not including pumps and UV) 24V dc Power consumption (peak demand): 2000VA Electrical protection rating 20 amps Noise level during recirculation <70dBA

Note: System requires approximately 4' x 6' of floor space for tanks, filtration Unit, and filters. -The floor drain must be a minimum of 3" and meet all applicable building codes and system requirements -Actual Unit measurement 32" Wide x35"Deep x72"High VA Chesapeake Community Based Outpatient Clinic (CBOC) Chesapeake HCC1 Page 10 of 10

Qty: 1 - Finishes/Options - Image PATHOLOGY AND LABORATORY LAB-C1012(1) Department, Room, Quantity

The contract, on USAspending

Federal procurement data the awarding office reported to FPDS, matched to this solicitation by its number.

UEI
EDQMKDZR8L13
CAGE
59UA0
Vendor location
Upland, CA
Contract
36C24624P0924, purchase order
Obligated
$31,455.00
Actions
1 between March 27, 2024 and March 27, 2024
Competition
Competed Under SAP, 3 offers received
Set-aside reported
No Set Aside Used.
Described as
Water Purification System
Match
solicitation number 36C24624Q0414 equals the FPDS solicitation identifier; same awarding office 36C246 (high confidence)

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. February 7, 2024

    Sources sought

    Due February 13, 2024 at 10:00 AM EST. SAM.gov, notice 50d80859ab0f459cb3a84498dcc2de5b

  2. February 29, 2024

    Presolicitation

    Due March 6, 2024 at 10:00 AM EST. SAM.gov, notice 01da7c306c6e4ab1b6ba3d64e13ecc93

  3. February 29, 2024

    Solicitation, originally presolicitation

    Due March 7, 2024 at 10:00 AM EST. SAM.gov, notice dd2a4a2c181246ccb2a61035c8eb1398