Skip to content
Abierto

Combined synopsis and solicitation

Portable Force Plate System-Brand Name and Equal (BNE)

RFQ-CDRH-2026-133635

Food and Drug Administration, FDA Office of Acq Grant SVCS. Other Measuring and Controlling Device Manufacturing.

Offers due

September 21, 2026 at 4:00 PM EDT

3 days from now. Posted September 17, 2026.

Description

As published on SAM.gov.

Notice of Combined Synopsis/Solicitation. This is a combined synopsis/solicitation for commercial products or commercial services prepared in accordance with part 12. This announcement constitutes the only solicitation; quotes are being requested, and a separate written solicitation will not be issued. The NAICS for this requirement is: 334519. The government is not obligated to make an award. The Government is not obligated to pay costs incurred by the Contractor during quote preparation and quote submission.

Background:

In collaboration with the Human-Device Interaction lab within the Office of Science and Engineering Labs (OSEL), the Division of

Biomedical Physics (DBP) seeks to order a portable force plate system for gait and balance assessment. The Human-Device

Interaction lab within OSEL addresses regulatory science research questions related to diagnostic, therapeutic, and assistive

medical devices. Understanding human interaction with these medical devices requires biomechanics equipment that quantifies

human movement. Force plate analysis is a common method by which human gait and balance can be accurately

quantified.

For this acquisition, a portable force plate system is needed to collect more robust, high-quality signals of human gait and balance.

The system will allow for accurate quantification of subject movement. The services requested below are for the operations of the

Center for Devices and Radiological Health (CDRH).
 

Minimum Performance Requirements/ Salient Characteristics:

The portable force plate system ordered must have the following salient characteristics:

o             A minimum of two (2) force plates capable of exporting data from at least six (6) measurement channels including force in the X, Y, and Z axes and moment in the X, Y, and Z axes. The following horizontal and vertical force and moment capacities must be met:

o             Force X, Y capacity must be greater than 430 Newtons.

o             Force Z capacity must be greater than 1300 Newtons.

o             Moment X, Y capacity must be greater than 225 Newton-meters.

o             Moment Z capacity must be greater than 80 Newton-meters.
o            Uses a plug & play Universal Serial Bus (USB) (2.0 or greater) interface capable of connecting and synchronizing up to 12 force plates. All cables and accessories required for synchronization should be included.

O          Generates  digital output of at least six measurement channels (described above) that uses a single USB cable to connect with an acquisition computer running Windows 10 and is easily integrated in Vicon Nexus 2.X software.

o             Compatible with Vicon Nexus software versions 2.6 through 2.19 (or latest available version)

o             Includes necessary drivers, plugins, or SDK components for integration

o             Real-time data streaming capability to Vicon Nexus environment
o            Includes at least 15-foot USB cable for power and data transfer

o          Force  plates allow for user selectable digital data rates between 10-1000 datasets per second.
o         Force plates do not require an external amplifier. Amplifier must be bult into the force plate.
o         Force plate dimensions are no greater than 51 x 51 x 4.6 centimeters with a 300-pound (or greater) vertical force capacity.
o         Force plate is portable such that it does not require any mounting and weighs less than 12 kilograms.
o         Hardware solutions are provided to recess the force plates into raised flooring such that data accuracy and integrity are not compromised.

O       The Contractor shall provide a minimum one (1) year hardware warranty from date of   Government acceptance. The Contractor shall guarantee full device replacement at no additional cost to the Government for any manufacturing defects or operational failures during normal use as defined in the user manual. Warranty coverage shall include all associated shipping costs for replacement units.

Place of Performance:

Hardware associated with the portable force plate system to be delivered no later than 60 days

after award.
 

SHIP TO:

Attn: Kimberly Kontson

10903 New Hampshire Ave

Building 62-1105

Silver Spring, MD 21140

Point of contact:

Kimberly Kontson

Kimberly.Kontson@fda.hhs.gov

Office phone: (301) 798-4990
 

Period of Performance:

12 consecutive months commencing upon date of award.

Quotes are due by email on or before September 21, 2026, by 4:00 P.M. and should be sent to brandon.rafus@fda.hhs.gov

Attachments

Files listed on SAM.gov. The ones we hold a copy of download from here; the rest open SAM's download, which needs no account.

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. September 17, 2026

    Combined synopsis and solicitation

    Due September 21, 2026 at 4:00 PM EDT. SAM.gov, notice 3d9a5cc3f46b422984a82c0edf3f55d3

    1 revisions on SAM
    • September 17, 2026: publish, combined synopsis and solicitation

Points of contact