VPRO Max2 Low Temp Sterilizers
36C25526Q0154
Department of Veterans Affairs, 255-Network Contract Office 15. Electromedical and Electrotherapeutic Apparatus Manufacturing.
Response deadline
January 21, 2026 at 12:00 PM EST
Closed 240 days ago. Posted January 14, 2026. Scheduled to archive February 5, 2026.
Description
As published on SAM.gov.
2. SPECIFICATIONS . Minimum of three low temperature sterilizers shall have a minimum individual dimensions of Cabinet (DD): 33 x 40-15/16 x 72” (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38- 5/16 x 72” (838 x 973 x 1829 mm). o Mounting single door, cabinet o System installation requires no plumbing, ventilation or air o supply – only a dedicated electrical connection is needed. A power cord is supplied for this connection o The facility space in which the vehicle can occupy is approx. 35’ x 110’. .
Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive materials because temperatures within the load currently do not exceed 55ºC (131ºF) and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized hydrogen peroxide sterilizers, either with or without plasma, are currently available in the United States. .
Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area with a ventilation system capable of at least 10 air exchanges per hour. . VPRO MaX2 must support utility requirements: o Current 16 AMP o Minimum Circuit Breaker 20 AMPS o Use NEMA L15-20R Plug o Provide grouped or ganged circuit protection and disconnect for phase 3 power as required by NEC and local codes. o Individual power shutoffs near each machine for servicing.
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
January 14, 2026
Sources sought
Due January 21, 2026 at 12:00 PM EST. SAM.gov, notice 4211ea88674841a19b166c31a5acd9f8
Points of contact
- Jon LutmerJonathan.lutmer@va.gov4104463862
Also open from this buyer
- 589A7-21-132 Construct Centralized Emergency Power SystemSolicitationSDVOSBNAICS 236220Wichita, KS36C25526R01250002Closes todaySep 18
- 36C25526Q0650 657-27-101, FCA-Conduct Hazardous Material Abatement, 2027 AE Design -FY27 NRMSources sought36C25526Q0650Closes todaySep 18
- Uninterruptible Power Supply St. Louis VA Medical CenterSources soughtNAICS 81121036C25526Q0678Closes todaySep 18
- Induction Charger for food warmerSources soughtSDVOSBNAICS 335999Kansas City, MO36C25526Q0657Closes todaySep 18
- BACKFLOW ANNUAL INSPECTION, TESTING & MAINTENANCE MARION, IL VA MEDICAL CENTERSources soughtSDVOSBNAICS 238220Illinois36C25527Q0005Closes todaySep 18