# VPRO Max2 Low Temp Sterilizers

Canonical: https://abierto.us/opportunities/36c25526q0154

- Solicitation number: 36C25526Q0154
- Notice type: Sources sought
- Status: Closed. Deadline was January 21, 2026 at 12:00 PM EST
- Department: Department of Veterans Affairs
- Contracting office: 255-Network Contract Office 15 (36C255)
- NAICS: 334510 Electromedical and Electrotherapeutic Apparatus Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Place of performance: Kansas City, Missouri
- County: Jackson County (FIPS 29095). https://abierto.us/counties/jackson-county-mo-29095
- City: Kansas City. https://abierto.us/cities/kansas-city-mo-2938000
- First posted: January 14, 2026
- Last posted: January 14, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/4211ea88674841a19b166c31a5acd9f8/view

## Description

2. SPECIFICATIONS . Minimum of three low temperature sterilizers shall have a minimum individual dimensions of Cabinet (DD): 33 x 40-15/16 x 72” (838 x 1040 x 1829 mm) with Recessed/Cabinet (SD): 33 x 38- 5/16 x 72” (838 x 973 x 1829 mm). o Mounting single door, cabinet o System installation requires no plumbing, ventilation or air o supply – only a dedicated electrical connection is needed. A power cord is supplied for this connection o The facility space in which the vehicle can occupy is approx. 35’ x 110’. .

Vaporized hydrogen peroxide sterilization processes are particularly suited for sterilizing heatsensitive materials because temperatures within the load currently do not exceed 55ºC (131ºF) and sterilization occurs in a low-moisture environment. Heat tolerant devices and materials can also be indicated for vaporized hydrogen peroxide sterilization processes. FDA-cleared vaporized hydrogen peroxide sterilizers, either with or without plasma, are currently available in the United States. .

Per ANSI/AAMI ST58:2013 or subsequent, the sterilizer shall be installed and operated in an area with a ventilation system capable of at least 10 air exchanges per hour. . VPRO MaX2 must support utility requirements: o Current 16 AMP o Minimum Circuit Breaker 20 AMPS o Use NEMA L15-20R Plug o Provide grouped or ganged circuit protection and disconnect for phase 3 power as required by NEC and local codes. o Individual power shutoffs near each machine for servicing.