4000 SYRINGE PUMP Name Brand or Equal To CS: Tracy Heath

36C26024Q0234

Department of Veterans Affairs, 260-Network Contract Office 20. Surgical and Medical Instrument Manufacturing.

Awarded

Marathon Medical Corporation

$34,504.14 obligated so far on USAspending

Description

As published on SAM.gov.

36C26024Q0234 0001 Description of Amendment: The purpose of this amendment is to:

update the salient characteristics for the software component. The two (2) bolded characteristics listed under software are not required and can be excluded. The quote due date is extended by one day and quotes are now due 02/13/2024 @ 15:00 PT. All other terms and conditions remain unchanged.

Salient Characteristics:

Infusion Pump: Accuracy +/- 2% Must be customizable with a design to help prevent medication errors and allow health care providers to send and receive medication delivery information efficiently. Capable of auto-detecting various syringe types and see, from 1mL to 60mL.

Dimensions: 6.2 H x 10.5 W x 5.7 D. Moisture protection or fluid resistant Tested to EMI/EMC testing per IEC 60601-2-24 Rechargeable battery: minimum 400 charge cycle counts, capable of 10 hour run time AC power: 100-240 VAC, 50/60 Hz, 42 VA; Safety Call II- Type CF with Functional Earth Conductor Warranty: one year factory warranty with a warranty extension available Operating conditions:

Temperature: 5 to 40 C (40 to 104 F)

Relative Humidity: 20 to 95% non-condensing Ambient Pressure: 70 kPa to 106 kPa (10.2 psia to 15 4 psia)

Storage and Transportation Conditions:

Temperature: -20 to 50 C (-4 to 122 F)

Relative Humidity: 20% to 95% non-condensing Ambient Pressure: 70 kPa to 106 kPa (10.2 psia to 15.4 psia)

Software: Pharmguard Medication Safety Software Comprehensize and versatile drug libraries Flexible workflows Designed to support the varying needs in specialized care areas. Reports highlighting important issues and allow for drill down in areas needing attention. Enhanced patient/clinician safety Streamlined clinical workflows with infusion data near real-time (auto transfer infusion data to EHR)

Rotating Pole Clamp: Rotating Pole clamp (45 and 90 degree)

The contract, on USAspending

Federal procurement data the awarding office reported to FPDS, matched to this solicitation by its number.

UEI
YNGKKQWQLKC9
CAGE
4KU99
Vendor location
Aurora, CO
Contract
36C26024P0330, purchase order
Obligated
$34,504.14
Actions
1 between February 21, 2024 and February 21, 2024
Competition
Competed Under SAP, 1 offer received
Set-aside reported
Small Business Set Aside - Total
Described as
4000 Syringe Pumps & Pharmguard Software
Match
solicitation number 36C26024Q0234 equals the FPDS solicitation identifier; same awarding office 36C260 (high confidence)

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. February 6, 2024

    Combined synopsis and solicitation

    4000 SYRINGE PUMP

    Due February 12, 2024 at 6:00 PM EST. SAM.gov, notice 21c2f3e4585949a88dc076d69b6e6053

  2. February 6, 2024

    Combined synopsis and solicitation

    02/29/2024 | 668 | $33,727.80 | 4000 SYRINGE PUMP

    Due February 12, 2024 at 6:00 PM EST. SAM.gov, notice 50e7244a45b2431081e6aadc937c121f

  3. February 6, 2024

    Combined synopsis and solicitation

    4000 SYRINGE PUMP

    Due February 12, 2024 at 6:00 PM EST. SAM.gov, notice 8feb9fcfa4874b6e893f88549d2fedb9

  4. February 12, 2024

    Combined synopsis and solicitation

    Due February 13, 2024 at 6:00 PM EST. SAM.gov, notice e28b9b1d400b4b999f983edcc1ad0210

Points of contact