Safety Air Monitoring and Analytical Referee System (SAMARS) Gas Chromatograph (GC)

W911S6-24-Q-0034

Department of the Army, W6QM Micc-Dugway Prov GRD. Analytical Laboratory Instrument Manufacturing.

Awarded

Teledyne Instruments Inc.

$228,674.00 on the award notice, March 24, 2025, contract W911S625P0002

Description

As published on SAM.gov.

Procure two GC instruments with the following specifications: Flame Photometric Detector (FPD); one instrument with phosphorous optical filter, one instrument with sulfur optical filter Chromatographic columns: one instrument DB-1701 phase and one instrument DB-210 phase column; Air sampling pre-concentrator trap capability with flow rate through the trap of at least 500 mL/min External mass flow controller, ability to capture the flow meter readings Oven programmable from ambient to 400?

Cycle time from injection to injection of 15 minutes or less Visible and audible alarms for system faults and sample concentration detections Quick change trap assembly, flow rate of 500 mL/min with ability to change the preconcentrator tube in approximately 1 minute Onboard integration engine with graphical use interface. Ability to operate the instrument, calibrate, troubleshoot, analyze samples, etc. from the user interface.

Visual readouts displayed of set points and measurements such as: sample flow, temperatures, pressures, concentrations, injection types, valve status, voltage, etc. Data logging. Ability to capture instrument set points and measured values such as: flow, temperatures, pressures, calibration, concentration, injection type – calibration, service, sample, etc. Data captured locally with the ability to retrieve and use for other reporting or calculating of sample data. Automatic calibration feature.

Calibration of the instrument using an averaged single point or multi-point calibration curve. Capable of calibrating and obtaining sample concentrations using the onboard integration engine, and values displayed on the graphical user interface. Heated exhaust to prevent expulsion of water. Detector exhaust must expel water in gas/vapor form elimination water drip from the instrument detector. Sample bypass capability. Ability for continuous sample flow at inlet in a bypass state. A minimum flow of 500 mL/min required for bypass capability. GC approximate dimensions - 17” x 11” x 8.5” (width x height x depth) Instrument weight approximately 25 pounds

The contract, on USAspending

Federal procurement data the awarding office reported to FPDS, matched to this solicitation by its number.

UEI
J1SQC3517JK5
CAGE
21070
Vendor location
City of Industry, CA
Contract
W911S625P0002, purchase order
Obligated
$228,674.00
Actions
3 between March 24, 2025 and August 19, 2025
Competition
Competed Under SAP, 3 offers received
Set-aside reported
No Set Aside Used.
Described as
Safety Air Monitoring and Analytical Referee System
Match
award number W911S625P0002 equals the contract number; same awarding office W911S6 (high confidence)

Publications

Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.

  1. February 25, 2025

    Combined synopsis and solicitation

    Due March 10, 2025 at 1:00 PM EDT. SAM.gov, notice 0b011432bab844789eb2e88e24a518b8

  2. April 8, 2025

    Award notice

    Awarded to TELEDYNE INSTRUMENTS Inc. City of Industry for $228,674. SAM.gov, notice 7902881e291d4c01a62da4541fee13f3