Combined synopsis and solicitation, total small business set-aside
Tokai Hit STXG-GSI2X stage top incubator for Leica Stellaris STED microscope
75N98026Q00794
National Institutes of Health, National Institutes of Health Olao. Analytical Laboratory Instrument Manufacturing.
Response deadline
July 25, 2026 at 9:00 AM EDT
Closed 53 days ago. Posted July 15, 2026. Scheduled to archive July 26, 2026.
Description
As published on SAM.gov.
Requirements: Provides accurate dynamic control of temperature, humidity, CO2 and oxygen to maintain required cell culture conditions for live cell confocal and STED super-resolution imaging. The stage top incubator is required in order to perform live cell imaging with the unique resolution capabilities of our Leica Stellaris 5 STED microscope. It currently does not have a stage top incubator, and therefore, is unable to perform this type of imaging requested by NCI researchers.
Stage top incubator controller must have two feedback modes, chamber temperature or specimen temperature, while it also continuously monitors the room temperature to maintain specimen temperature within set point anywhere between 25 – 50 degrees Celsius. The built-in digital gas mixer controller accepts 100% CO2 from lab supply to regulate CO2 concentration at the sample, with a CO2 range of 5.0 to 20% with real-time sample feedback.
In addition, humidity is actively controlled with an internal humidifier to sense sample chamber humidity. The electrically heated stage top incubator chamber must have magnets for easy interchange of specimen holders, including inserts for 35 mm coverglass dish, 50/60mm coverglass dish, chamber slide, coverglass chamber, and open frame insert for imaging multi-well chambers. A flexible heated collar that accommodates a range of objective lens diameters must also be included. The components of the stage top incubator must include:
Tokai Hit STXG-GSI2X-SET Stage Top Incubator for Super Z galvo stage (inverted microscope stands) Pure (100%CO2) gas use (110- 240V) for individual temperature settings to create optimum temperature and humidity conditions without difficulty for long term live cell imaging. Setting temperature range from 25 to 50°C.
Dimensions of 112 x 87 mm compatible with Leica Super Z galvo stage on existing Leica Stellaris 5 STED microscope. the lightest chamber (less than 140g) for Super Z Glavo stage temperature controller with built-in digital gas mixer (CO2: 5.0 - 20.0%) Objective heater Magnetic Dish holder and fixing lid for 35mm, 50/60mm dish, chamber slide, slide glass and chamber slide use. External Sensor, enable to autoclave repeatability for Sample Feedback regulation. Tokai Hit original data logging, temp/CO2 program control, software compatible to use with LAS-X software
Publications
Every notice SAM.gov issued under this solicitation number, oldest first. Each is a separate record on SAM.
July 15, 2026
Combined synopsis and solicitation
Due July 25, 2026 at 9:00 AM EDT. SAM.gov, notice 7f0309dced09447697a2aca2279013b4
Points of contact
- John Burtonburtonjo@mail.nih.gov
Also open from this buyer
- Acquisition of Confocal Microscope Laser System ReCombined synopsis and solicitationNAICS 33451675N98026Q01163Closes todaySep 16
- Cryo-Electron Microscopy (Cryo-EM) ServicesCombined synopsis and solicitationNAICS 54171475N98026Q01177Closes todaySep 16
- Nikon ORCA-Quest 2 cameraCombined synopsis and solicitationNAICS 33451675N98026Q01159Closes todaySep 16
- Upgrade for LUMICKS C-trap to 3 color confocal instrumentCombined synopsis and solicitationNAICS 811210North Carolina75N98026Q01151-04750Closes todaySep 16
- Specialized Clinical Laboratory Testing ServicesPresolicitationNAICS 541380Maryland75N98026Q01207Closes tomorrowSep 17