# Purchase of One (1) QS3 qPCR System, One (1) iBright FL1500 Imaging System, and Associated Accessories, Brand Name or Equal To for Oklahoma City VA Medical Center (OKCVAMC)

Canonical: https://abierto.us/opportunities/36c25927q0016

- Solicitation number: 36C25927Q0016
- Notice type: Sources sought
- Status: Open. Offers due October 15, 2026 at 11:00 AM EDT
- Department: Department of Veterans Affairs
- Contracting office: Network Contract Office 19 (36C259)
- NAICS: 334516 Analytical Laboratory Instrument Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Place of performance: Oklahoma City, Oklahoma
- Who can bid, by location: Anywhere in the US. Location does not limit who can bid: a product shipped to the buyer, research at the performer's facility, or work at the contractor's site. A state shown is the delivery point.
- County: Oklahoma County (FIPS 40109). https://abierto.us/counties/oklahoma-county-ok-40109
- City: Oklahoma City. https://abierto.us/cities/oklahoma-city-ok-4055000
- First posted: October 8, 2026
- Last posted: October 8, 2026
- SAM.gov: https://sam.gov/workspace/contract/opp/80f7b6a3bcad4b9ca8613356afa8f213/view

## Description

Sources Sought Notice   Purchase of One (1) QS3 qPCR System, One (1) iBright FL1500 Imaging System, and Associated Accessories, Brand Name or Equal To for Oklahoma City VA Medical Center (OKCVAMC)
Disclaimer:
This Request for Information (RFI) is issued solely for market research information and planning purposes and does not constitute a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. Responses to this notice are NOT offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.

Information Requested from Industry:

In response to the Sources Sought, interested Contractors shall submit the following information and any Capabilities/Qualifications Statement to include but not limited to an overview of proposed solution(s) and a description of the product your company possesses in accordance with the Statement of Work draft / Salient Characteristics List.

Responses to market research table questionnaire shall be submitted. If this information is not provided, then it will be assumed the entire requirement cannot be met.

Department of Veterans Affairs, Veterans Health Administration (VHA), Network Contracting Office 19 (NCO 19) is conducting a market survey and is seeking potential sources for Purchase of One (1) QS3 qPCR System, One (1) iBright FL1500 Imaging System, and Associated Accessories, Brand Name or Equal To for Oklahoma City VA Medical Center (OKCVAMC) for Oklahoma City VA Medical Center (OKCVAMC). This Sources Sought Notice is issued for information and planning purposes only. This is not a solicitation or a request for proposal and shall not be construed as an obligation or commitment by the Government. An award will not be made on any offers submitted in response to this notice, and it shall not be implied the Government is committed to providing any solicitation or award following this notice. The Government will not pay for any information received in response to this request, nor will the Government compensate a respondent for any costs incurred in developing the information provided.

Brand Name or Equal to QS3 qPCR System - Qty 1. The offer must meet the capabilities of the line items listed in addition to the following salient characteristics:
Description
Part Number
Qty
QS3 0.1 mL qPCR System with Laptop
A28566
1
Fast 96 Well RNase P Verification Plate
4351979
1
Fast 96 Well Hardshell Clear Plate (20 pack)
4483485
1
Optical Adhesive Covers (25 per pack)
4360954
1
802.11ac/b/g/n Wi Fi Dongle
A26774
1
Fast 96 Well ROI & Background Plates
4432426
1
PowerTrack SYBR Master Mix, 1 mL
A46012
1
TaqMan Fast Advanced Master Mix, 1 mL
4444556
1
SmartStart qStudio 3 Orientation
A28806
1

Salient Characteristics:
System and Consumables
QuantStudio 3 Real-Time PCR System
Compatible with Applied Biosystems MicroAmp EnduraPlate optical 96-well fast clear reaction plates
Supports barcoded, 0.1 mL reaction plates
Includes or identifies the quantity of plates provided
Identifies any additional consumables required for operation
Thermal Control and Block Configuration
96-well sample block
At least three independently controlled Peltier temperature zones
Programmable temperature difference of up to 10Â°C across the block
Maximum temperature difference of 5Â°C between adjacent zones
Block temperature range of approximately 4Â°C to 99.9Â°C
Reaction Volumes and Thermal Performance
Supports reaction volumes of approximately 10 30 ÂµL on a 0.1 mL block
Supports reaction volumes of approximately 10 100 ÂµL on a 0.2 mL block
Maximum ramp rate of approximately 9.0Â°C per second for the 0.1 mL block
Maximum ramp rate of approximately 6.5Â°C per second for the 0.2 mL block
Optical Detection and Heated Cover
Long-life white LED light source with a median operating life of at least 60,000 hours
CMOS imaging system for data acquisition
Data collection for each filter combination in less than two seconds
Heated cover temperature range of approximately 50Â°C to 110Â°C
Heated cover designed to minimize evaporation from reaction plates
Storage, Connectivity, Power, and Reliability
Compatible with the Connect platform
Provides up to 1 TB of cloud-based storage
Includes approximately 10 GB of onboard storage
Stores approximately 1,000 5,000 standard absolute-quantification run files
Compatible with 100 240 V electrical outlets without modification
Provides run protection during network or power interruptions
Includes information on data security, software requirements, warranty, technical support, maintenance, training, and recurring storage or software costs

Brand Name or Equal to iBright FL1500 Imaging System - Qty 1. The offer must meet the capabilities of the line items listed in addition to the following salient characteristics:
Description
Part Number
Qty
iBright FL1500 Imaging System
A44241
1
SmartStart iBright Orientation
A34016
1
1 Preventive Maintenance visit for iBright FL1500
ZG61SCIB
1
Bolt Welcome Pack with iBlot 3
NW0412AI
1
iBright Protein Standard (25 ÂµL)
LC5605
1
iBright FL1500 2 Year Warranty

1

Salient Characteristics:
Imaging Capabilities and Performance
Chemiluminescent western blot imaging
Protein gel, DNA gel, membrane, and stained SDS-PAGE gel imaging
Automatic image capture, focus, exposure, and analysis
Quantitative band analysis and background subtraction
High sensitivity, broad dynamic range, and low-light detection
Imaging of multiple samples, including up to four mini gels in one view
Camera, Detection, Illumination, and Imaging Area
Cooled scientific-grade CCD camera
At least 16-bit image capture and approximately 9-megapixel resolution or higher
Adjustable pixel binning and optical/digital zoom
Imaging area of approximately 22.5 cm by 18 cm or larger
Removable sample tray
White-light and blue/green-light illumination
Optional near-infrared illumination
Chemiluminescent detection, multiple filter options, and long-life LED light sources
Interface, Software, Storage, and Connectivity
Integrated touchscreen and internal computer or workstation
Automated focus, exposure, zoom, image enhancement, quantification, annotation, reporting, and analysis
Molecular-weight analysis, where applicable
TIFF, JPEG, PNG, PDF, and other common image formats
At least 250 GB of internal storage
USB and Ethernet connectivity
Optional wireless or cloud connectivity
Secure data transfer, user access controls, and software updates

Accessories and Training
Sample tray, imaging accessories, and white screen or equivalent for stained-gel imaging
User manuals, quick-start guides, installation instructions, and training materials
3. Scope:
The Oklahoma City VA Medical Center requires the Applied Biosystems QS3 qPCR System, iBright FL1500 Imaging System, and associated accessories to support research into osteoarthritis treatment and pain management.

4. Delivery:
Department of Veterans Affairs
Oklahoma City VA Medical Center
921 Northeast 13th Street
Oklahoma City, OK 73104-5007

In response to this announcement, please provide the information requested below requested in this market research table questionnaire. If this information is not provided, then it will be assumed the entire requirement cannot be met:
NAICS Code: 334516
Company Name:
Address:
UEI (Unique Entity ID) Number:
Contact Name:
Phone No.:
Email:
Business Size Information - Select all that applies:
Small Business
Emerging Small Business
Small Disadvantaged Business
Certified under Section 8(a) of the Small Business Act
HUBZone
Woman Owned
Certified Service-Disabled Veteran Owned Small
Veteran Owned Small Business
Large Business
FSS/GSA Contract Holder:
Yes                     No
FSS/GSA Contract Number:

Effective Date/ Expiration Date:

Proposed solution is listed and available on the above FSS/GSA Contract:
Yes                     No
Pricing Model
Please Indicate Availability Below: (Yes / No / NA)
All on FSS
Open Market only
Mix of FSS & Open Market
Firm-Fixed Price

Federal Acquisition Regulation (FAR) Market Rearch Questions:

Buy American Act RFO 52.225-1 and RFO 52.225-2 Buy American Certificate   What percentage of the proposed product (including leases and rentals) is a:

Domestic end product? _____________ (%)

Foreign end product? _______________ (%) Country of Origin:

Questions for Small Businesses ONLY:

If subcontracting, what added value do you offer (RFO 52.215-23):

Nonmanufacturer Rule (RFO 52.219-33):
Does your company manufacturer these proposed items? [Â  ] yes [Â  ] no
Does your company exceed 500 employees?  [Â  ] yes [Â  ] no   If yes, list # of employees: _________
Does your company primarily engaged in the retail or wholesale trade and normally sells the type of item being supplied?  [Â  ] yes [Â  ] no
Does your company take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice?  [Â  ] yes [Â  ] no
Does your company supply the end item of a small business manufacturer, processor or producer made in the United States, or obtains a waiver of such requirement pursuant to paragraph (b)(5) CFR 121.406. [Â  ] yes [Â  ] no  If yes, what is the manufacturer s name? ________________

Response Instructions:
Do not provide more than 8 pages, including cover letter page.
Submit your response via email to devin.reichert@va.gov
Submit your response NLT 09:00 (MST) on October, 15th, 2026;
Mark your response as  Proprietary Information  if the information is considered business sensitive.
VA has identified the appropriate North American Industry Classification System (NAICS) Code as 334516 (Analytical Laboratory Instrument Manufacturing) with a Small Business size of 1000 Employees.  Please identify and explain any NAICS codes your company believes would better represent the predominant work included in the attached Statement of Work draft / Salient Characteristics List.

(End of Document)

## Attachments

- [36C25927Q0016.docx](https://sam.gov/api/prod/opps/v3/opportunities/resources/files/dd6aeeb1835541f7b24ec0fb98170b33/download)

## Open now in the same industry

30 open in NAICS 334516.

- Fiber Coupled MgOLN Phase Modulator (6-B187-Q-00459-00), due October 12, 2026 at 10:00 AM EDT. https://abierto.us/opportunities/6b187q0045900
- Linear X-Nanopositioners (6-B187-Q-00463-00), due October 12, 2026 at 1:30 PM EDT. https://abierto.us/opportunities/6b187q0046300
- LT-SPM (BSA475591), due October 12, 2026 at 4:00 PM EDT. https://abierto.us/opportunities/bsa475591

## Publications

- October 8, 2026: Sources sought, due October 15, 2026 at 11:00 AM EDT. Notice 80f7b6a3bcad4b9ca8613356afa8f213. https://sam.gov/workspace/contract/opp/80f7b6a3bcad4b9ca8613356afa8f213/view

## Points of contact

- Devin Reichert, Contract Specialist, Devin.Reichert@va.gov, 303-712-5709

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://abierto.us/opportunities/36c25927q0016.