# Medical Grade Electrometer for Oncology testing

Canonical: https://abierto.us/opportunities/36c25925q0746

- Solicitation number: 36C25925Q0746
- Notice type: Sources sought
- Status: Closed. Deadline was September 3, 2025 at 11:00 AM EDT
- Department: Department of Veterans Affairs
- Contracting office: Network Contract Office 19 (36C259)
- NAICS: 339113 Surgical Appliance and Supplies Manufacturing
- Product or service code: 6515 Medical and Surgical Instruments, Equipment, and Supplies
- Place of performance: Oklahoma City, Oklahoma
- County: Oklahoma County (FIPS 40109). https://abierto.us/counties/oklahoma-county-ok-40109
- City: Oklahoma City. https://abierto.us/cities/oklahoma-city-ok-4055000
- First posted: August 20, 2025
- Last posted: August 20, 2025
- SAM.gov: https://sam.gov/workspace/contract/opp/141a5314f4384ffab20fd4e5e02dcbf2/view

## Description

The Oklahoma City Veterans Affairs Health Care System is seeking sources for a medical electrometer to perform tests performed by the Radiation Oncology department.

The device is anticipated to have the following requirements: Minimum requirements Remote Access Full remote access through a web-browser Flash compatible Supports your TG-51 QA protocols Connectivity via WiFi, Ethernet or USB 3.0 Host Correction Factors Full remote access through a web-browser Flash compatible Supports your TG-51 QA protocols Connectivity via WiFi, Ethernet or USB 3.0 Host Library Functions Extended detector library Check source library Machine library API Functionality Large touch screen Desired specifications Connectivity Ethernet LAN 1 Gbit/s, Wi-Fi (802.11ac/b/g/n), USB 3.0 Host Dimensions (at or near) 249.7mm Ã— 201.9mm Ã— 123.4mm Current Range 400 fA 24 A Charge Range From 24 pC to 65 nC per pulse at a repetition frequency of 400 Hz Configurations TNC Triax, BNC Triax (Adapter) Zeroing 60 s (default) Bias Voltage Â?} 50 V Â?} 500 V, Â?} 2 % Zero Drift Â?} 0.25 % ( Â?} 0.5 %, IEC 60731) Non-linearity Â?} 0.25 % ( Â?} 0.5 %, IEC 60731) Response Time < 1.5 s Long Term Stability Â?} 0.2 % ( Â?} 0.5 % over one year, IEC 60731) Repeatability < Â?} 0.25 % ( Â?} 0.25 % IEC 60731) Measurement Modes Manual, Automatic (beam triggered), Timed Resolution 0.1 fC Additional information This is a request for information and does not represent an order or any obligation on behalf of the VA nor does it commit the VA to a future request for quote (RFQ).

Please provide the country of manufacturing for devices you would consider quoting if an RFQ is issued. If you consider your company a potential source for this type of device please provide the following information: Company name, Name of a point of contact including email address, Size demographics of company Questions? Deadline for questions 8/26 Point of contact: Christopher.delauro@va.gov

## Publications

- August 20, 2025: Sources sought, due September 3, 2025 at 11:00 AM EDT. Notice 141a5314f4384ffab20fd4e5e02dcbf2. https://sam.gov/workspace/contract/opp/141a5314f4384ffab20fd4e5e02dcbf2/view

## Points of contact

- christopher.delauro@va.gov, Chris DeLauro, christopher.delauro@va.gov, 303-712-5737

---
Source: SAM.gov Contract Opportunities bulk extract. Confirm deadlines on SAM.gov before responding. Cite https://abierto.us/opportunities/36c25925q0746.